MEDTRONIC EXTENDED
Report
- Report Number
- 8021545-2024-02992
- Event Type
- Malfunction
- Date Received
- August 6, 2024
- Date of Event
- July 4, 2024
- Report Date
- December 17, 2024
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244022805
- PMA / PMN Number
- K210544
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
E1: (B)(6).
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. CAPA-2010953 "TREND OBSERVED: INCREASE IN COMPLAINTS RELATED TO ADHESIVE ON EWIS" HAS BEEN OPENED ON 18-SEP-2024 TO ADDRESS ALL ADHESIVE ISSUES RELATED TO EWIS PRODUCT FAMILY AS NO SPECIFICATIONS EXIST FOR THE ADHESIVE USED ON THIS PRODUCT (DESIGN RELATED CAPA).
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY "TREND OBSERVED: INCREASE IN COMPLAINTS RELATED TO ADHESIVE ON EWIS" HAS BEEN OPENED ON 18-SEP-2024 TO ADDRESS ALL ADHESIVE ISSUES RELATED TO EWIS PRODUCT FAMILY AS NO SPECIFICATIONS EXIST FOR THE ADHESIVE USED ON THIS PRODUCT (DESIGN RELATED CAPA).
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT, PATIENT FACED INFUSION SET TAPE NOT STICKING EVENT ON 04-JUL-2024. THE EVENT OCCURED AFTER THE PATIENT WENT FOR SWIMMING. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1674602 | MEDTRONIC EXTENDED | UNO EWIS BLUE 60/9 HCAP 10PK INT | FPA | UNOMEDICAL A/S | MMT-441A | 6005372 | 05705244022805 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |