FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 1991594 · Received January 28, 2011

Report

Report Number
2032227-2011-00292
Event Type
Malfunction
Date Received
January 28, 2011
Date of Event
September 15, 2010
Report Date
January 26, 2010
Manufacturer
MEDTRONIC MINIMED
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD A FROZEN DISPLAY DUE TO CORROSION ON THE LIQUID CRYSTAL DISPLAY BOARD. NO DAMAGE WAS FOUND ON THE KEYPAD TRACE. UNABLE TO PERFORM THE OPERATING CURRENTS DUE TO THE FROZEN DISPLAY.

Description of Event or Problem · 1

THE CUSTOMER CALLED TO REPORT UNRESPONSIVE BUTTONS FOR OVER AN HOUR. THE INSULIN PUMP WAS NOT DROPPED OR BUMPED. ADVISED THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC MINIMED MMT-722LNAP

Patients

Seq Age Sex Outcome Treatment
1 24 YR