Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP ALARMED DURING THE DISPLACEMENT TEST AND ALARMED MOTOR ERROR DUE TO MOTOR EXCESSIVE AXIAL PLAY.
Description of Event or Problem · 1
THE CUSTOMER REPORTED A MOTOR ERROR ALARM DURING THE MANUAL PRIME. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. ADVISED THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.