FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 1991569 · Received January 28, 2011

Report

Report Number
2032227-2011-00296
Event Type
Malfunction
Date Received
January 28, 2011
Date of Event
January 17, 2011
Report Date
January 17, 2011
Manufacturer
MEDTRONIC MINIMED
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED DURING THE DISPLACEMENT TEST AND ALARMED MOTOR ERROR DUE TO MOTOR EXCESSIVE AXIAL PLAY.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A MOTOR ERROR ALARM DURING THE MANUAL PRIME. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. ADVISED THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC MINIMED MMT-522LNAP

Patients

Seq Age Sex Outcome Treatment
1 28 YR