FDA Adverse Event Injury Summary report: N

CAPSURE SP NOVUS

MDR report key: 1991376 · Received February 15, 2011

Report

Report Number
2649622-2011-02367
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P850089/S38
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD IS OVERSENSING IN THE BIPOLAR MODE, POSSIBLY DUE TO A PAST MICROLEAD DISLODGMENT. THE LEAD REMAINS IN USE AND NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSURE SP NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5092 ASKU

Patients

Seq Age Sex Outcome Treatment
1 88 YR Required Intervention SEDR01 IMPLANTABLE PULSE GENERATOR| 5594 IMPLANTABLE PACING LEAD