FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1991342 · Received February 15, 2011

Report

Report Number
2649622-2011-02361
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, THE OUTER INSULATION WAS BREACHED CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THE HELIX/LOBE WAS DISTORTED/BENT AND THERE WAS APPARENT EXPLANT DAMAGE. EVALUATION SUMMARY FOR (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND THE DISTAL CONDUCTOR FRACTURED. IT WAS NOTED THAT THE INNER INSULATION WAS KINKED/BUCKLED, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THE LEAD WAS STRETCHED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL LEAD HAD A POSSIBLE FRACTURE AND THE RIGHT VENTRICULAR LEAD HAD HIGH PACING THRESHOLDS. BOTH LEADS WERE REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 38 YR Hospitalization| R ADDR01 IMPLANTABLE PULSE GENERATOR