FDA Adverse Event Malfunction Summary report: N

PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED

MDR report key: 1991333 · Received February 15, 2011

Report

Report Number
1423500-2011-01975
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
February 2, 2011
Report Date
February 2, 2011
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K053512
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED OPERATIONAL AND IN GOOD CONDITION. THE DEVICE PASSED THE HOMECHOICE RITE (RETURN INSTRUMENT TEST / EVALUATION) FUNCTIONAL TEST PASSED BUT FAILED THE RITE ELECTRICAL GROUND BOND TEST. THE PAL EVALUATED THE DEVICE. THE GROUND BUS RESISTANCE WAS MEASURED AND FOUND TO BE WITHIN THE GROUND BOND SPECIFICATION. THE ASSIGNABLE CAUSE FOR RITE TEST FAILURE - GROUND BOND WAS UNDETERMINED. THE DEVICE'S SERVICE HISTORY RECORD WAS REVIEWED AND NO ISSUES WERE IDENTIFIED THAT MAY HAVE CONTRIBUTED TO THE RITE FAILURE. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, A BAXTER TECHNICIAN DETERMINED THE HC MACHINE FAILED THE GROUND BOND TEST INDICATING A HIGH RESISTANCE VALUE BETWEEN THE HC AND THE GROUND BOND ON THE POWER SUPPLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PD CYCLER 110 VOLT HOME CHOICEPRO AUTOMATED SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1