FDA Adverse Event Injury Summary report: N

ATTAIN OTW

MDR report key: 1991327 · Received February 15, 2011

Report

Report Number
2649622-2011-02353
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P010015/S003
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED AND ANALYZED. PRIMARY RESULTS REVEALED THERE WERE NO ANOMALIES FOUND. BLOOD/BODY FLUID WAS PRESENT ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED). THE OUTER INSULATION WAS MELTED, HAD COSMETIC ENVIRONMENTAL STRESS CRACK (ESC) AND COSMETIC DEPRESSION. THERE WAS ALSO APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR LEAD DISLODGED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN OTW IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4193 ASKU

Patients

Seq Age Sex Outcome Treatment
1 82 YR Hospitalization| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6949 IMPLANTABLE TACHY LEAD