FDA Adverse Event Injury Summary report: N

SIGMA 200 DR

MDR report key: 1991293 · Received February 15, 2011

Report

Report Number
6000094-2011-00259
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC S.A.
Product Code
DXY
PMA / PMN Number
P980035/S2
Removal / Correction Number
Z-1510/1519-2009
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) PRELIMINARY ANALYSIS REVEALED NO OUTPUT AND NO TELEMETRY. FURTHER TESTING REVEALED THAT THE NO OUTPUT AND NO TELEMETRY CONDITIONS WERE THE RESULT OF LIFTED HYBRID BOND WIRES. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE COULD NOT BE INTERROGATED. ON ELECTROGRAM ( ECG), IT WAS REPORTED THERE WAS JUST PATIENT INTRINSIC RHYTHM, NO STIMULATION BY THE PACEMAKER. THE DEVICE WAS EXTRACTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SIGMA 200 DR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC S.A. SDR203 ASKU

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R