FDA Adverse Event Malfunction Summary report: N

ATTAIN OTW

MDR report key: 1991266 · Received February 15, 2011

Report

Report Number
2649622-2011-02314
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
December 17, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P010015/S012
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED).

Description of Event or Problem · 1

IT WAS REPORTED THAT, DURING AN IMPLANT ATTEMPT, THE PATIENT ANATOMY WAS TORTUOUS AND THE LEAD WAS DIFFICULT TO POSITION/FIX AND THE THRESHOLDS WERE UNACCEPTABLE. A DIFFERENT LEAD WAS SELECTED FOR PLACEMENT IN A SMALLER VEIN AND WAS IMPLANTED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ATTAIN OTW IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4194 ASKU

Patients

Seq Age Sex Outcome Treatment
1 57 YR Other