PRECICE STRYDE SYSTEM
Report
- Report Number
- 3006179046-2024-00156
- Event Type
- Malfunction
- Date Received
- August 5, 2024
- Date of Event
- August 12, 2020
- Report Date
- July 31, 2024
- Manufacturer
- NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED
- Product Code
- HSB
- UDI-DI
- 00887517951304
- PMA / PMN Number
- K180198
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
THE DEVICE WAS RETURNED TO THE LONDON IMPLANT RETRIEVAL CENTRE (LIRC); NO PRODUCT WAS RETURNED TO THE MANUFACTURER FOR INVESTIGATION. ALL PRECICE STRYDE DEVICES HAVE BEEN RECALLED SINCE FEBRUARY 2021 AND ARE CURRENTLY OFF THE MARKET. A CAPA WAS COMPLETED TO ADDRESS THE REPORTED EVENT. THE CAPA FOUND THE RISK MANAGEMENT (RM) PROCESS WAS INADEQUATE. FUTURE RM FILES FOR ALL ITERATIONS OF THE DEVICE WILL BE INTEGRATED FROM THE NUVASIVE SPECIALIZED ORTHOPEDICS-SPECIFIC RISK MANAGEMENT AND DESIGN CONTROL PROCESS/PROCEDURES INTO THE NUVASIVE, INC. GLOBAL RISK MANAGEMENT PROCESS/PROCEDURE. A REVIEW OF THE DHR DOCUMENTS INDICATES THE DEVICE WAS MANUFACTURED BY THE SPECIFIED REQUIREMENT AT THE TIME AND MET ALL THE REQUIRED INSPECTIONS UPON RELEASE TO FINISHED GOODS.
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED, DETAILING THE EXAMINATION OF THE RETRIEVED IMPLANT. DURING THE ASSESSMENT, IT WAS NOTED THAT THE IMPLANT WAS UNABLE TO BE DISTRACTED AND RETRACTED AND FAILED THE FORCE TESTING. ADDITIONALLY, SURFACE DAMAGE GRADING REVEALED NOTABLE CORROSION, SCRATCHING, GALLING, SURFACE DEPOSITS AND DISCOLORATION. THE REPORT INDICATED THAT THE ROD WAS REMOVED AS PART OF A PLANNED PROCEDURE FOLLOWING THE ACHIEVEMENT OF THE DESIRED LENGTHENING. THE PATIENT HAD PREVIOUSLY REPORTED EXPERIENCING MILD THIGH PAIN, AND RADIOGRAPHIC OBSERVATIONS IDENTIFIED CORTICAL THICKENING AT THE EXTENDABLE JUNCTION OF THE NAIL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2252686 | PRECICE STRYDE SYSTEM | ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES | HSB | NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED | PS13.0-50D235 | 8092403 | 00887517951304 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Female |