CLARITY DDDR
Report
- Report Number
- 6000094-2011-00251
- Event Type
- Injury
- Date Received
- February 15, 2011
- Manufacturer
- MEDTRONIC S.A.
- Product Code
- DXY
- PMA / PMN Number
- P990001/S4
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED AND ANALYZED. THE BATTERY DEPLETION WAS NORMAL.
IT WAS REPORTED THAT THE DEVICE HAD A POWER ON RESET AND SUSPECTED PREMATURE BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. IT WAS ALSO REPORTED THAT THE DEVICE COULD NOT BE INTERROGATED SUCCESSFULLY. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THE DEVICE HAD A POWER ON RESET AND SUSPECTED PREMATURE BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CLARITY DDDR | IMPLANTABLE PULSE GENERATOR | DXY | MEDTRONIC S.A. | 860 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Hospitalization| R |