FDA Adverse Event Injury Summary report: N

ENRHYTHM DR

MDR report key: 1991152 · Received February 15, 2011

Report

Report Number
6000144-2011-00838
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
DXY
PMA / PMN Number
P980035/S38
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) THE ACTUAL DEVICE WAS RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. BATTERY VOLTAGE - BATTERY DEPLETION INDICATED/ERI ERI DUE TO HIGH BATTERY IMPEDANCE WAS TRIGGERED ON (B)(6) 2010, PRIOR TO EXPLANT. RAMWARE VERSION 8 WAS INSTALLED ON (B)(6) 2010. ANALYSIS OF DEVICE FOUND BATTERY HAD HIGH IMPEDANCE. THE INCORRECT RTT BATTERY VOLTAGE MEASUREMENTS ARE THE RESULT OF HIGH INTERNAL BATTERY RESISTANCE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE CLINIC SYMPTOMATIC. THE DEVICE BATTERY WAS AT ELECTIVE REPLACEMENT INDICATION, AND THE BATTERY WAS ALLEGED TO HAVE PREMATURELY DEPLETED. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENRHYTHM DR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC MED REL, INC. P1501DR ASKU

Patients

Seq Age Sex Outcome Treatment
1 78 YR Hospitalization| R 4076X2 IMPLANTABLE PACING LEAD| 4076X2 IMPLANTABLE PACING LEAD| 4076X1 IMPLANTABLE PACING LEAD