FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II Y 24GAX0.75IN PRN/EC SLM

MDR report key: 19911405 · Received August 5, 2024

Report

Report Number
3002601200-2024-00371
Event Type
Malfunction
Date Received
August 5, 2024
Date of Event
June 24, 2024
Report Date
July 18, 2024
Manufacturer
BD SUZHOU (MDS)
Product Code
FOZ
UDI-DI
00382903830336
PMA / PMN Number
UNK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

IN RESPONSE TO THE EVENT REPORTED BY YOUR FACILITY A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 3080088. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, VISUAL ANALYSIS WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RETAINED DEVICES WERE FOUND TO BE FREE OF ANY DAMAGE OR OTHER ABNORMALITIES. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT. BD WILL CONTINUE TO MONITOR THIS ISSUE AND ENCOURAGES YOU TO SUBMIT YOUR SAMPLE FOR REVIEW.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INTIMA-II Y 24GAX0.75IN PRN/EC SLM HAD FOREIGN MATTER. (B)(6) 2024 AT 08:30 HOURS, PATIENT WAS ADMITTED TO THE HOSPITAL WITH A COLONIC MALIGNANCY FOR A PCT, AND WHEN THE NURSE PLACED A TUBE IN HER, SHE USED AN AIRTIGHT IV NEEDLE, WHICH RESULTED IN A GRADE III ADVERSE EVENT THAT DID NOT CAUSE ANY HARM TO THE PATIENT. FOREIGN BODY CONTAMINATION OF THE NEEDLE OF THE CLOSED INDWELLING NEEDLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2301292 BD INTIMA-II Y 24GAX0.75IN PRN/EC SLM INTRAVASCULAR CATHETER FOZ BD SUZHOU (MDS) 3080088 00382903830336

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown