FDA Adverse Event Injury Summary report: N

SCREW-IN

MDR report key: 1991135 · Received February 15, 2011

Report

Report Number
2649622-2011-02263
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
K902002
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. FOR MODEL 5568. EVALUATION SUMMARY: (B)(4) NO ANOMALIES FOUND. ADDITIONAL FINDINGS INCLUDE PROXIMAL CONDUCTOR DISTORTED, DISTAL CONDUCTOR FRACTURE (OVERSTRESS), OUTER INSULATION BREACHED CUT AND COSMETIC DEPRESSION, AND APPARENT EXPLANT DAMAGE. PARTIAL LEAD IN SEGMENTS RETURNED AND ANALYZED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL LEAD HAD LOSS OF CAPTURE, HIGH IMPEDANCE, AND WAS APPARENTLY DISLODGED. THE RIGHT ATRIAL LEAD WAS EXPLANTED AND REPLACED. IT WAS FURTHER REPORTED THAT THE LEFT VENTRICULAR LEAD HAD HIGH THRESHOLDS. THE LEFT VENTRICULAR LEAD IS STILL IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SCREW-IN IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5071 ASKU

Patients

Seq Age Sex Outcome Treatment
1 78 YR Hospitalization| R 7299 IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 7299 IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6949 IMPLANTABLE TACHY LEAD| 6949 IMPLANTABLE TACHY LEAD