FDA Adverse Event
Injury
Summary report: N
SELECT SECURE
MDR report key: 1991082
·
Received February 15, 2011
Report
- Report Number
- 2649622-2011-02248
- Event Type
- Injury
- Date Received
- February 15, 2011
- Date of Event
- September 8, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- NVN
- PMA / PMN Number
- P030036
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE HELIX/LOBE WAS DISTORTED/BENT AND THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEAD HAD NO CAPTURE, UNDERSENSING AND DIFFICULTY WITH POSITIONING/FIXATION. THE LEAD WAS DISLODGED. IT WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SELECT SECURE | IMPLANTABLE PACING LEAD | NVN | MEDTRONIC PUERTO RICO, INC. | 3830 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Hospitalization| R | (B)(4) IMPLANTABLE PULSE GENERATOR| 3830 IMPLANTABLE PACING LEAD |