INFINITI VISION SYSTEM OZIL
Report
- Report Number
- 2028159-2011-00085
- Event Type
- Injury
- Date Received
- February 10, 2011
- Date of Event
- January 7, 2011
- Report Date
- January 10, 2011
- Manufacturer
- ALCON - IRVINE TECHNOLOGY CENTER
- Product Code
- HQC
- PMA / PMN Number
- K082845
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
NO FIELD SERVICE WAS REQUESTED AND NO SAMPLES ARE EXPECTED TO RETURN FOR IN-HOUSE EVALUATION. A ROOT CAUSE HAS NOT BEEN IDENTIFIED. CORNEAL BURN IS AN ISSUE THAT IS OCCASIONALLY REPORTED WITH CATARACT SURGERY. ACCORDING TO THE (B)(6) PT SAFETY ADVISORY ABSTRACT: PREVENTING CORNEAL BURNS DURING PHACOEMULSIFICATION, MARCH 2010, VOL. 7, NO. 1: 23-25, MOST CORNEAL BURNS CAN BE TRACED TO ISSUES RELATED TO SURGICAL TECHNIQUE AND NOT TO MALFUNCTIONING EQUIPMENT. A LETTER AND COPY OF THIS INDUSTRY ARTICLE WAS PROVIDED TO THE CUSTOMER. (B)(4).
A SURGEON REPORTED A CORNEAL BURN OCCURRED AT THE END OF THE PHACOEMULSIFICATION PHASE OF A MODEST NUCLEUS EXTRACTION PROCEDURE. A SMALL WHITE PATCH WAS OBSERVED AT THE PROXIMAL SIDE OF THE INCISION THAT MADE CLOSURE VERY DIFFICULT. TWO SUTURES WERE REQUIRED. IN A FOLLOW UP, THE SURGEON REPORTED THE PHACO TIME WAS NOT LONG AND THE CHOP PROCEDURE WAS USED. THE PT IS DOING WELL. ADDITIONAL INFO HAS BEEN REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFINITI VISION SYSTEM OZIL | PHACOFRAGMENTATION SYSTEM | HQC | ALCON - IRVINE TECHNOLOGY CENTER | INFINITI OZIL | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | PROVISC |