FDA Adverse Event Injury Summary report: N

ENRHYTHM DR

MDR report key: 1990842 · Received February 15, 2011

Report

Report Number
6000144-2011-00782
Event Type
Injury
Date Received
February 15, 2011
Date of Event
December 5, 2010
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
DXY
PMA / PMN Number
P980035/S38
Removal / Correction Number
Z-1440-2010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4): THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA WHICH REVEALED THAT THE DEVICE REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) ON (B)(6) 2010.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED AND ANALYZED. CALCULATIONS BASED ON IMPLANT PARAMETERS INDICATE THAT THE DEVICE HAD MET ITS 99.9% EXPECTED LONGEVITY. ANALYSIS OF THE DEVICE MEMORY FOUND THE ERI (ELECTIVE REPLACEMENT INDICATOR) IS THE RESULT OF HIGH INTERNAL BATTERY RESISTANCE. THIS DEVICE IS PART OF THE FIELD ACTION AND HAS TESTED CONSISTENT WITH THE FIELD ACTION. PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA WHICH REVEALED THAT THE DEVICE REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) ON (B)(6) 2010. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4): THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA WHICH REVEALED THAT THE DEVICE REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) ON (B)(6) 2010. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE REACHED ELECTIVE REPLACEMENT INDICATOR AFTER NEW SOFTWARE WAS LOADED. THE DEVICE WAS REPROGRAMMED. IT WAS FURTHER REPORTED THAT THE DEVICE WAS AT EARLY ELECTIVE REPLACEMENT INDICATOR. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE REACHED ELECTIVE REPLACEMENT INDICATOR AFTER NEW SOFTWARE WAS LOADED. THE DEVICE WAS REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENRHYTHM DR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC MED REL, INC. P1501DR ASKU

Patients

Seq Age Sex Outcome Treatment
1 77 YR Hospitalization| R 5076 X2 IMPLANTABLE PACING LEAD| 5076X2 IMPLANTABLE PACING LEAD| 5076 X2 IMPLANTABLE PACING LEAD