FDA Adverse Event Injury Summary report: N

CAPSUREFIX

MDR report key: 1990824 · Received February 15, 2011

Report

Report Number
2649622-2011-02141
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S2
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD WAS CAPPED DUE TO A FRACTURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5068 ASKU

Patients

Seq Age Sex Outcome Treatment
1 94 YR Hospitalization| R (B)(4) IMPLANTABLE PULSE GENERATOR| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 5524M IMPLANTABLE PACING LEAD