FDA Adverse Event Injury Summary report: N

CAPSUREFIX

MDR report key: 1990818 · Received February 15, 2011

Report

Report Number
2649622-2011-02139
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S2
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION WAS PULLED APART DUE TO OVERSTRESS, THE OUTER INSULATION HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, THE OUTER INSULATION WAS BREACHED CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THERE WAS APPARENT EXPLANT DAMAGE AND THERE WAS A WHITE SUBSTANCE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL LEAD HAD A DECREASE IN IMPEDANCE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4568 ASKU

Patients

Seq Age Sex Outcome Treatment
1 83 YR Hospitalization| R (B)(4) IMPLANTABLE PULSE GENERATOR| 4068 IMPLANTABLE PACING LEAD