FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1990808 · Received February 15, 2011

Report

Report Number
2649622-2011-02133
Event Type
Injury
Date Received
February 15, 2011
Date of Event
December 1, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
Z-0475-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY : (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED; THE HELIX WAS DISENGAGED FROM HELICAL CHANNEL. THE DISTAL CONDUCTOR WAS DISTORTED, THERE WAS BLOOD IN/ON THE HELIX MECHANISM SLEEVE HEAD AND ON THE HELIX MECHANISM ITSELF, AND THERE WAS A TIP SEAL OBSERVATION. THERE WAS A DEFORMATION/FRACTURE IN THE PROXIMAL SECTION OF THE INNER COIL, INDICATING EXTENSION PROBLEMS. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD HAD HIGH THRESHOLDS. UPON LEAD REVISION, THE HELIX WOULD NOT RETRACT. THE LEAD WAS REMOVED AND REPLACED. THERE ARE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 84 YR Hospitalization| R 5076 IMPLANTABLE PACING LEAD| 4196 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 6947 IMPLANTABLE TACHY LEAD| 4196 IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB