FDA Adverse Event Injury Summary report: N

EON MINI IPG, 16-CHANNEL RECHARGEABLE

MDR report key: 1990713 · Received February 8, 2011

Report

Report Number
1627487-2011-00182
Event Type
Injury
Date Received
February 8, 2011
Date of Event
January 14, 2011
Report Date
January 14, 2011
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION DIVISION
Product Code
LGW
PMA / PMN Number
P010032
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL: METHOD: THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS: REVIEW OF THE DEVICE HISTORY RECORD FOUND A NONCONFORMANCE; HOWEVER, THE NONCONFORMANCE WAS IDENTIFIED AS A COSMETIC ISSUE AND DID NOT AFFECT PRODUCT INTEGRITY OR PRODUCT FUNCTIONALITY AND WAS APPROVED FOR USE. THE DHR ANOMALY IS NOT RELATED TO THE ALLEGED DEVICE FAILURE. CONCLUSION: THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. SJM HAS LIMITED INFO RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REFERENCE MFR REPORT # 1627487-2011-00181. THE PT RECEIVED HIS SCS SYSTEM ON (B)(6) 2010 CONSISTING OF IPG AND SURGICAL LEAD. IT WAS REPORTED THAT HE DEVELOPED AN INFECTION. THERE ARE NO PLANS FOR SURGICAL INTERVENTION AT THIS TIME AS THE PHYSICIAN HAS ELECTED TO LEAVE THE SYSTEM IMPLANTED. ORAL ANTIBIOTICS WERE ADMINISTERED TO THE PT AS TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EON MINI IPG, 16-CHANNEL RECHARGEABLE TOTALLY IMPLANTABLE PULSE GENERATOR LGW ST. JUDE MEDICAL - NEUROMODULATION DIVISION 3228 3117475

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention