FDA Adverse Event Malfunction Summary report: N

GDC-10 360 DEGREE COIL SR; GDC-10 360 DEGREE SOFT COIL SR; GDC-18 360 DEGREE COI

MDR report key: 1990664 · Received February 15, 2011

Report

Report Number
2939204-2011-00039
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
December 5, 2010
Report Date
January 24, 2011
Manufacturer
BOSTON SCIENTIFIC - CORK
Product Code
HCG
PMA / PMN Number
K042539
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATION. VISUAL INSPECTION IDENTIFIED SMALL BLOOD MATTER IN THE DISTAL END OF THE INTRODUCER SHEATH SUGGESTIVE THAT INSUFFICIENT FLUSH HAD BEEN MAINTAINED. HOWEVER, THERE WAS ONLY A SMALL AMOUNT OF BLOOD AND THIS COULD HAVE BEEN DUE TO BACK FLOW DURING REMOVAL OF THE DEVICE. THE PROXIMAL END OF THE MAIN COIL WAS PROTRUDING FROM THE PROXIMAL END OF THE INTRODUCER SHEATH. THE PUSHER WIRE WAS KINKED 9CM FROM THE DISTAL END. THE PROXIMAL END THE COIL WAS EXTENSIVELY STRETCHED AND THE MID TO DISTAL END WAS KINKED. MICROSCOPE ANALYSIS OF THE COIL REVEALED THAT THE PET JUNCTION WAS BROKEN. THE INNER COIL WAS STILL PRESENT ON THE DISTAL END OF THE PUSHER WIRE INDICATING THAT THE COIL HAD NOT DETACHED BY ELECTROLYSIS AS THE DETACHMENT ZONE IS INTACT. IT IS LIKELY THAT THE COIL STRETCHED AS IT WAS ATTEMPTED TO WITHDRAW THE DEVICE AND IT IS PROBABLE THAT THE DEVICE BROKE AT THE PET JUNCTION DUE TO EXCESSIVE FORCE APPLIED TO THE DEVICE IN THE ATTEMPT TO REMOVE IT. THEREFORE, A ROOT CAUSE OF OPERATIONAL CONTEXT WILL BE ASSIGNED TO THIS INVESTIGATION.

Description of Event or Problem · 1

DEVICE ANALYSIS FOUND THAT THE POLYETHYLENE (PET) JUNCTION WAS BROKEN. THERE WAS NO REPORTED CLINICAL CONSEQUENCE TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GDC-10 360 DEGREE COIL SR; GDC-10 360 DEGREE SOFT COIL SR; GDC-18 360 DEGREE COI DEVICE, NEUROVASCULAR EMBOLIZATION HCG BOSTON SCIENTIFIC - CORK M0033461030SR0 12560079

Patients

Seq Age Sex Outcome Treatment
1 48 YR