FDA Adverse Event Injury Summary report: N

BINAXNOW

MDR report key: 1990658 · Received February 8, 2011

Report

Report Number
1221359-2011-00001
Event Type
Injury
Date Received
February 8, 2011
Date of Event
January 10, 2011
Report Date
February 8, 2011
Manufacturer
ALERE SCARBOROUGH, INC.
Product Code
GNX
PMA / PMN Number
K053126
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL LOT# 2363. ADDITIONAL EXP DATE: 11/30/2014. THE NASOPHARYNGEAL SWAB IS INCLUDED AS A COMPONENT OF THE BINAXNOW INFLUENZA A & B TEST KIT. THE SWAB IS A PACKAGED COMPONENT PURCHASED FORM (B)(4). PURITAN WAS CONTACTED AND INFORMED OF THIS ADVERSE EVENT. THE SAMPLE WAS RETURNED TO PURITAN MEDICAL PRODUCTS (B)(4), MANUFACTURER OF THE SWAB. THE FOLLOWING WAS RECEIVED FROM PURITAN. (B)(4). BATCH RECORDS WERE REVIEWED AS WELL AS RETENTION SAMPLES PULLED FOR EVALUATION. NO DEFECTIVE OR ANOMALIES WERE DISCOVERED. BASED ON SAMPLES EXAMINED, SHAFTS DO NOT FRACTURE UNLESS BENT REPEATEDLY. IT IS BELIEVED THE END USER MANIPULATED THE SWAB PRIOR TO IT ENTERING THE NASAL PASSAGE. (B)(4). IT IS DETERMINED THAT THIS IS AN ISOLATED EVENT GIVEN THE HISTORICAL EVIDENCE OF THE CLAIM, THE EVALUATION OF THE SAMPLES. THIS IS AN EXTRAORDINARILY RARE INCIDENCE.

Description of Event or Problem · 1

WHEN COLLECTION A NASOPHARYNGEAL SAMPLE FOR THE BINAXNOW INFLUENZA A&B TEST KIT, THE PLASTIC SHAFT OF THE COLLECTION SWAB PROVIDED IN THE KIT BROKE WHERE THE FOAM AND SHAFT MEET. THE PLASTIC SHAFT AND FOAM WERE LODGED IN THE PT'S NASAL PASSAGE, AND THE DOCTOR NOTED THAT THE PT SNIFFED WHILE THE SAMPLE WAS BEING COLLECTED. THE PARENTS REFUSED ANY INTERVENTION FOR REMOVAL OF TIP. THEY WERE NOT INTERESTED IN GOING TO THE EMERGENCY ROOM. THE DOCTOR FOLLOWED UP WITH THE PARENTS ON (B)(6) 2011 AND (B)(6) 2011 AND THE CHILD WAS NOT UNCOMFORTABLE AND THE PARENTS STILL DO NOT WANT ANY FURTHER INVESTIGATIONS PERFORMED ON CHILD TO RETRIEVE SWAB TIP. THE DOCTOR IS HOPING THAT THE CHILD SWALLOWED THE TIP AND PASSED IT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BINAXNOW INFLUENZA A&B CLIA WAIVED 22 TEST KIT GNX ALERE SCARBOROUGH, INC. 042827 OR 043170

Patients

Seq Age Sex Outcome Treatment
1 3 YR Other