FDA Adverse Event Malfunction Summary report: N

ENDO GIA

MDR report key: 19906525 · Received August 5, 2024

Report

Report Number
2647580-2024-03425
Event Type
Malfunction
Date Received
August 5, 2024
Date of Event
May 30, 2024
Report Date
August 23, 2024
Manufacturer
US SURGICAL PUERTO RICO
Product Code
GDW
UDI-DI
10884523003215
PMA / PMN Number
K111825
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3 EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. VISUAL INSPECTION NOTED THE CLAMPING MECHANISM WAS DEFORMED AND EVALUATION NOTED THAT THE INTERLOCK WAS OVERRIDDEN. IT WAS REPORTED THAT THE INSTRUMENT WAS DIFFICULT TO UNLOAD. THE REPORTED ISSUE WAS CONFIRMED. THE PRODUCT ANALYSIS NOTED EVIDENCE THAT THE DEVICE WAS NOT USED AS INTENDED. THIS ISSUE CAN OCCUR UNDER THE FOLLOWING CONDITIONS. APPLICATION OVER TISSUE THAT IS BEYOND THE RECOMMENDED THICKNESS RANGE. APPLICATION WITH AN OBSTACLE INCORPORATED IN THE JAWS. IN ANY OF THESE CIRCUMSTANCES, IT WILL BECOME INCREASINGLY DIFFICULT TO ACTUATE THE FIRING HANDLE AND THE INSTRUMENT RETURN KNOBS WILL BE DIFFICULT TO RETRACT. IN ADDITION, STAPLES MAY NOT FORM PROPERLY AND TISSUE MAY NOT BE FULLY TRANSECTED. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. THE INSTRUCTIONS INCLUDED WITH THIS DEVICE PROVIDE THE FOLLOWING GUIDANCE: FAILURE TO COMPLETELY FIRE THE RELOAD WILL RESULT IN AN INCOMPLETE CUT AND/OR INCOMPLETE STAPLE FORMATION, WHICH MAY RESULT IN POOR HEMOSTASIS AND/OR LEAKAGE. ALWAYS INSPECT THE TISSUE THICKNESS AND SELECT AN APPROPRIATE STAPLE SIZE PRIOR TO APPLICATION OF THE ENDO GIA¿ ULTRA UNIVERSAL SHORT, ENDO GIA¿ ULTRA UNIVERSAL OR ENDO GIA¿ ULTRA UNIVERSAL XL STAPLER. OVERLY THICK OR THIN TISSUE MAY RESULT IN UNACCEPTABLE STAPLE FORMATION. WHEN POSITIONING THE STAPLER ON THE APPLICATION SITE, ENSURE THAT NO OBSTRUCTIONS, SUCH AS CLIPS, ARE INCORPORATED INTO THE INSTRUMENT JAWS. FIRING OVER AN OBSTRUCTION MAY RESULT IN INCOMPLETE CUTTING ACTION AND/OR IMPROPERLY FORMED STAPLES. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

D10 CONCOMITANT PRODUCTS: EGIAUSTND - EGIAUSTND ENDOGIA ULTRA UNIV STD STAP, LOT# P2E0208S MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

ACCORDING TO THE REPORTER, DURING A THORACOSCOPIC LOBECTOMY PROCEDURE, WHILE DETACHING LUNG LOBE TISSUE, THE HANDLE AND RELOAD COULD NOT BE DISASSEMBLED. A NEW HANDLE AND RELOAD WERE USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY. PLI:20 MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT DEVICE FOUND THAT THE CLAMPING MECHANISM WAS DEFORMED DUE TO THICK TISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1723324 ENDO GIA STAPLE, IMPLANTABLE GDW US SURGICAL PUERTO RICO EGIA60AMT P2G0098S 10884523003215

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown