FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1990601 · Received February 15, 2011

Report

Report Number
2649622-2011-02087
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
November 19, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND THE DISTAL CONDUCTOR FRACTURED DUE TO OVERSTRESS. IT WAS NOTED THAT THE DISTAL CONDUCTOR WAS DISTORTED, THE HELIX/LOBE WAS DISTORTED/BENT, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THE LEAD APPEARED DAMAGED AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ATTEMPTED IMPLANT THE HELIX WAS UNABLE TO ADVANCE AND RETRACT. THE LEAD WAS NOT USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 64 YR Other