ENRHYTHM DR
Report
- Report Number
- 6000094-2011-00232
- Event Type
- Injury
- Date Received
- February 15, 2011
- Manufacturer
- MEDTRONIC S.A.
- Product Code
- DXY
- PMA / PMN Number
- P980035/S38
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GR
- Reporter Occupation
- PHYSICIAN
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED AND ANALYSIS FOUND THE BATTERY IMPEDANCE WAS HIGH WHICH TRIGGERED THE ELECTIVE REPLACEMENT INDICATOR. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS.
IT WAS REPORTED THAT THE DEVICE'S ELECTIVE REPLACEMENT INDICATOR WAS TRIGGERED AND THERE WAS POSSIBLE PREMATURE BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
ASKU
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENRHYTHM DR | IMPLANTABLE PULSE GENERATOR | DXY | MEDTRONIC S.A. | P1501DR | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R | 5054 IMPLANTABLE PACING LEAD| 5554 IMPLANTABLE PACING LEAD |