SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2011-02069
- Event Type
- Injury
- Date Received
- February 15, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) HELIX DISENGAGED FROM HELICAL CHANNEL. ADDITIONAL FINDINGS OF DEFIB CONDUCTOR DISTORTED AND BLOOD/BODY FLUID (NOT OBSTRUCTED, INNER TUBING KINKED/BUCKLED, OUTER TUBING OVERLAY COSMETIC ENVIRONMENTAL STRESS CRACKING, ALL INSULATORS BREACHED CUT, OUTER INSULATION COSMETIC DEPRESSION AND BLOOD IN/ON HELIX MECHANISM. PARTIAL LEAD IN SEGMENTS RETURNED AND ANALYZED.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD HAD A LOSS OF CAPTURE AND A GRADUAL INCREASE IN THRESHOLD. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Hospitalization| R | (B)(4) COMPETITOR IMPLANTABLE PULSE GENERATOR| (B)(4) COMPETITOR IMPLANTABLE PULSE GENERATOR |