FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1990522 · Received February 15, 2011

Report

Report Number
2649622-2011-02047
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD IN SEGMENTS WAS RETURNED AND ANALYZED. OUTER INSULATION BREACHED DEPRESSION. THE PROXIMAL CONDUCTOR HAD BLOOD/BODY FLUID (NOT OBSTRUCTED); OUTER INSULATION HAD COSMETIC DEPRESSION; BLOOD IN/ON THE HELIX/LOBE MECHANISM.

Description of Event or Problem · 1

INFECTION WAS REPORTED. THE LEAD WAS REMOVED, ANALYZED AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 70 YR Hospitalization| R IMPLANTABLE PULSE GENERATOR| 5076 IMPLANTABLE PACING LEAD