FDA Adverse Event Malfunction Summary report: N

CAPSURE SP NOVUS

MDR report key: 1990500 · Received February 15, 2011

Report

Report Number
2649622-2011-02030
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
December 8, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P850089/S38
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY : (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED WITH NO ANOMALIES FOUND. THE OUTER INSULATION HAD A COSMETIC CUT AND THERE WAS APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN IMPLANT ATTEMPT, THE RIGHT VENTRICULAR LEAD HAD POSITIONING/FIXATION DIFFICULTIES. THE LEAD WAS NOT USED, RATHER IT WAS EXCHANGED FOR AN ACTIVE FIXATION LEAD WHICH WAS IMPLANTED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSURE SP NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5092 ASKU

Patients

Seq Age Sex Outcome Treatment
1 70 YR Other