CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2011-02029
- Event Type
- Malfunction
- Date Received
- February 15, 2011
- Date of Event
- December 1, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S017
- Removal / Correction Number
- ASKU
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM.
IT WAS REPORTED THAT DURING A POST OPERATION CHECK THE RIGHT VENTRICULAR (RV) LEAD HAD HIGH THRESHOLDS. A LEAD REVISION WAS ATTEMPTED, BUT REVEALED HIGH IMPEDANCE WHEN REPOSITIONED. THE LEAD WAS REMOVED AND A SECOND (RV) LEAD WAS ATTEMPTED. THE SECOND RV LEAD ALSO SHOWED HIGH IMPEDANCE AND THE HELIX RETRACTED THEN COULD NOT REDEPLOY. THE SECOND RV LEAD WAS ALSO NOT USED. A THIRD RV LEAD WAS IMPLANTED WITH GOOD RESULTS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 4076 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Other | (B)(4) IMPLANTABLE PULSE GENERATOR| 4076 IMPLANTABLE PACING LEAD |