FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1990499 · Received February 15, 2011

Report

Report Number
2649622-2011-02029
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
December 1, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S017
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A POST OPERATION CHECK THE RIGHT VENTRICULAR (RV) LEAD HAD HIGH THRESHOLDS. A LEAD REVISION WAS ATTEMPTED, BUT REVEALED HIGH IMPEDANCE WHEN REPOSITIONED. THE LEAD WAS REMOVED AND A SECOND (RV) LEAD WAS ATTEMPTED. THE SECOND RV LEAD ALSO SHOWED HIGH IMPEDANCE AND THE HELIX RETRACTED THEN COULD NOT REDEPLOY. THE SECOND RV LEAD WAS ALSO NOT USED. A THIRD RV LEAD WAS IMPLANTED WITH GOOD RESULTS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 80 YR Other (B)(4) IMPLANTABLE PULSE GENERATOR| 4076 IMPLANTABLE PACING LEAD