FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 1990445 · Received February 15, 2011

Report

Report Number
2649622-2011-01997
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED) AND THE OUTER INSULATION WAS BREACHED CUT. IT WAS NOTED THAT THE STEROID RING WAS DAMAGED DURING ANALYSIS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTED LEAD BECAME DISLODGED AND A NEW LEAD WAS ATTEMPTED TO BE IMPLANTED. THE SECOND LEAD WAS NOT IMPLANTED AS NOISE WAS OBSERVED ON THE ANALYZER. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 76 YR Hospitalization| R COMPETITOR IMPLANTABLE PACING LEAD| COMPETITOR IMPLANTABLE PULSE GENERATOR