CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2011-01997
- Event Type
- Injury
- Date Received
- February 15, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED AND THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED) AND THE OUTER INSULATION WAS BREACHED CUT. IT WAS NOTED THAT THE STEROID RING WAS DAMAGED DURING ANALYSIS.
IT WAS REPORTED THAT THE IMPLANTED LEAD BECAME DISLODGED AND A NEW LEAD WAS ATTEMPTED TO BE IMPLANTED. THE SECOND LEAD WAS NOT IMPLANTED AS NOISE WAS OBSERVED ON THE ANALYZER. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Hospitalization| R | COMPETITOR IMPLANTABLE PACING LEAD| COMPETITOR IMPLANTABLE PULSE GENERATOR |