SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2011-01975
- Event Type
- Malfunction
- Date Received
- February 15, 2011
- Date of Event
- December 9, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- Z-0475-2011
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY : (B)(4): THE FULL LEAD WAS RETURNED AND ANALYZED; THE HELIX WAS DISENGAGED FROM HELICAL CHANNEL. THERE WAS BLOOD/BODY FLUID (NOT OBSTRUCTED) ON THE DISTAL CONDUCTOR AND OUTER TUBING OVERLAY. THE OUTER TUBING OVERLAY HAD A BREACHED CUT, THERE WAS BLOOD IN/ON THE HELIX MECHANISM, AND THERE WAS A TIP SEAL OBSERVATION. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE.
IT WAS REPORTED THAT THE SMALL R-WAVES WERE NOTED AND THERE WAS AND ATTEMPT TO MOVE THE LEAD, SUBSEQUENTLY THE "LEAD WAS RETRACTED 1 TIME ONLY 12 TURNS AND NOW CANNOT BE EXTENDED.". IT WAS ALSO REPORTED THAT THERE WAS AN INSULATION BREECH DUE TO USING THE LEAD TO GET A NEW GUIDE WIRE DOWN. THE LEAD WAS NOT IMPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
ASKU
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 86 YR | Other |