FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 1990410 · Received February 15, 2011

Report

Report Number
2649622-2011-01971
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S17
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICLE AND SUPERIOR VENA CAVA IMPEDANCES HAVE SPIKED TO 150 OHMS. THE PHYSICIAN DECIDED TO EXTRACT THE LEAD. DEVICE IS STILL IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6944 ASKU

Patients

Seq Age Sex Outcome Treatment
1 63 YR Hospitalization| R 7278 IMPLANTABLE PACEMAKER/CARDIO/DEFIB| COMPETITOR IMPLANTABLE PACING LEAD| 5076 IMPLANTABLE PACING LEAD