FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1990401 · Received February 15, 2011

Report

Report Number
2649622-2011-01966
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
December 9, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE FULL LEAD WAS RETURNED, ANALYZED, AND THE HELIX HAD DISENGAGED FROM HELICAL CHANNEL. IT WAS ALSO NOTED THAT THE INNER TUBING WAS KINKED/BUCKLED AND THAT THERE WAS BLOOD IN/ON THE HELIX MECHANISM.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEAD HELIX WOULD NOT EXTEND AFTER BEING RETRACTED. THE LEAD WAS ATTEMPTED AND NOT USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 79 YR Other (B)(4) ANNULOPLASTY RING| (B)(4) ANNULOPLASTY RING