FDA Adverse Event Injury Summary report: N

ENRHYTHM DR

MDR report key: 1990371 · Received February 15, 2011

Report

Report Number
6000144-2011-00714
Event Type
Injury
Date Received
February 15, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
DXY
PMA / PMN Number
P980035/S38
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED WHERE IT WAS DISCOVERED THAT THE BATTERY HAD HIGH IMPEDANCE, WHICH RESULTED IN EARLY RECOMMENDED REPLACEMENT TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE LASTED LESS THAN FIVE YEARS. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENRHYTHM DR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC MED REL, INC. P1501DR ASKU

Patients

Seq Age Sex Outcome Treatment
1 71 YR Hospitalization| R 5076 IMPLANTABLE PACING LEAD| 5568 IMPLANTABLE PACING LEAD