AUTOSOFT 90
Report
- Report Number
- 3003442380-2024-18853
- Event Type
- Malfunction
- Date Received
- August 3, 2024
- Date of Event
- May 25, 2024
- Report Date
- August 5, 2024
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- UDI-DI
- 05705244018303
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
SUPPLEMENTAL REPORT 01 - MDR 1931396 - MDR 3003442380-2024-18853. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H11: INVESTIGATION SUMMARY: SINCE NO LOT NUMBER IS AVAILABLE. A DETAILED INVESTIGATION, TESTS ON REFERENCE SAMPLES OR BATCH REVIEW CANNOT BE CONDUCTED. THEREFORE, THIS COMPLAINT WILL BE CLOSED, THIS ISSUE WILL BE MONITORED THROUGH POSTMARKETING SURVEILLANCE (PMS) PRODUCT TRENDS AND MALFUNCTION. IF ANY TRENDS PICKED UP, THIS WOULD FLAG ON THE TRIPS AND ALERTS ACCORDING TO THE ON MARKET QUALITY REVIEW (OMQR) PROCEDURE.
INITIAL AND FINAL MDR (B)(4).
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED TEN KINKED CANNULA EVENTS ON 25-MAY-2024. THE EVENT OCCURRED AFTER THREE HOURS OF INSERTION. THE INFUSION SET WAS IN USE FOR 1 DAY. THE INFUSION SET WAS INSERTED IN ABDOMEN. BLOOD GLUCOSE LEVELS DURING THE EVENT WAS HIGH. PATIENT ADDRESS HIGH BLOOD GLUCOSE LEVEL VIA TAKING MULTI DAILY INJECTION. PATIENT REPLACED INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY, DO NOT SEE KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION WAS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 951707 | AUTOSOFT 90 | UNO INSET II 60/6 BLUE TCAP 10PK INT | FPA | UNOMEDICAL A/S | 1002823 | UNKNOWN | 05705244018303 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 27 YR | Female |