FDA Adverse Event Malfunction Summary report: N

AUTOSOFT 90

MDR report key: 19902253 · Received August 3, 2024

Report

Report Number
3003442380-2024-18851
Event Type
Malfunction
Date Received
August 3, 2024
Date of Event
May 25, 2024
Report Date
August 5, 2024
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244018303
PMA / PMN Number
K032854
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT 01 - MDR 1931396. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H11: INVESTIGATION SUMMARY SINCE NO LOT NUMBER IS AVAILABLE. A DETAILED INVESTIGATION, TESTS ON REFERENCE SAMPLES OR BATCH REVIEW CANNOT BE CONDUCTED. THEREFORE, THIS COMPLAINT WILL BE CLOSED, THIS ISSUE WILL BE MONITORED THROUGH POSTMARKETING SURVEILLANCE (PMS) PRODUCT TRENDS AND MALFUNCTION. IF ANY TRENDS PICKED UP, THIS WOULD FLAG ON THE TRIPS AND ALERTS ACCORDING TO THE ON MARKET QUALITY REVIEW (OMQR) PROCEDURE.

Additional Manufacturer Narrative · 0

INITIAL AND FINAL MDR (B)(4). DEVICE 4 OF 10

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED TEN KINKED CANNULA EVENTS ON 25-MAY-2024. THE EVENT OCCURRED AFTER THREE HOURS OF INSERTION. THE INFUSION SET WAS IN USE FOR 1 DAY. THE INFUSION SET WAS INSERTED IN ABDOMEN. BLOOD GLUCOSE LEVELS DURING THE EVENT WAS HIGH. PATIENT ADDRESS HIGH BLOOD GLUCOSE LEVEL VIA TAKING MULTI DAILY INJECTION. PATIENT REPLACED INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY, DO NOT SEE KINKING AS BEING RELATED TO HUMAN FACTORS, BUT RATHER AS A TRAINING ISSUE INCLUDING CORRECT CHOICES OF INSERTION SITES AND INFUSION SETS AND CANNULA LENGTH. FURTHERMORE, THE SOFT CANNULA IS A FLEXIBLE MATERIAL THAT DURING USE AND UPON REMOVAL CAN BEND SLIGHTLY. NO FURTHER INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
951704 AUTOSOFT 90 UNO INSET II 60/6 BLUE TCAP 10PK INT FPA UNOMEDICAL A/S 1002823 UNKNOWN 05705244018303

Patients

Seq Age Sex Outcome Treatment
1 27 YR Female