CAPSUREFIX NOVUS
Report
- Report Number
- 2649622-2011-01889
- Event Type
- Injury
- Date Received
- February 15, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- DTB
- PMA / PMN Number
- P930039/S009
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4): THE FULL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE PROXIMAL CONDUCTOR WAS STRETCHED, THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACHED CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THERE WAS APPARENT EXPLANT DAMAGE.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY (B)(4): THE FULL LEAD WAS RETURNED IN SEGMENTS, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT THE PROXIMAL CONDUCTOR WAS STRETCHED, THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACHED CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THERE WAS BLOOD IN/ON THE HELIX MECHANISM AND THERE WAS APPARENT EXPLANT DAMAGE. (B)(4) THE DEVICE WAS RETURNED AND ANALYZED. THE BATTERY IMPEDANCE WAS FOUND TO BE HIGH.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD EXHIBITED HIGH IMPEDANCE PROBLEMS AND HIGH THRESHOLD. THE LEAD WAS REMOVED AND REPLACED WITH A NEW LEAD. THE DEVICE WAS EXPLANTED, RETURNED AND ANALYZED. THE DEVICE TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD EXHIBITED HIGH IMPEDANCE PROBLEMS AND HIGH THRESHOLD. THE LEAD WAS REMOVED AND REPLACED WITH A NEW LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSUREFIX NOVUS | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC PUERTO RICO, INC. | 5076 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 85 YR | Hospitalization| R | 5076 IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PULSE GENERATOR| 5076 IMPLANTABLE PACING LEAD |