FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 1990136 · Received February 15, 2011

Report

Report Number
2649622-2011-01846
Event Type
Injury
Date Received
February 15, 2011
Date of Event
December 1, 2010
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
LWS
PMA / PMN Number
P920015/S24
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE ACTUAL LEAD WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. THREE - VENTRICULAR NST (NON-SUSTAINED TACHYCARDIA) <=210 MS AVERAGE V-CYCLE BETWEEN (B)(6) 2010 07:24:51 AND (B)(6) 2010 07:19:32. TWO - VF=190 MS ON (B)(6) 2010 12:21:50 AND ON (B)(6) 2010 08:28:09. VENTRICULAR SHORT INTERVAL COUNT V-SIC=97.4 COUNTS AVG/DAY, IN 2.10 DAYS, BETWEEN (B)(6) 2010 06:02:31 AND (B)(6) 2010 08:33:40.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD PRODUCED NOISE AND OVERSENSING. AFTER FURTHER INVESTIGATION, IT WAS DETERMINED THE LEAD HAD A FRACTURE. THE LEAD IS NO LONGER IN SERVICE AND WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD PRODUCED NOISE AND OVERSENSING. AFTER FURTHER INVESTIGATION, IT WAS DETERMINED THE LEAD HAD A FRACTURE. THE LEAD IS NO LONGER IN SERVICE AND WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPRINT QUATTRO SECURE IMPLANTABLE TACHY LEAD LWS MEDTRONIC PUERTO RICO, INC. 6947 ASKU

Patients

Seq Age Sex Outcome Treatment
1 68 YR Hospitalization| O| R (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4439 COMPETITOR IMPLANTABLE PACING LEAD| 4439 COMPETITOR IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB