SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2011-01846
- Event Type
- Injury
- Date Received
- February 15, 2011
- Date of Event
- December 1, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- P920015/S24
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE ACTUAL LEAD WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. THREE - VENTRICULAR NST (NON-SUSTAINED TACHYCARDIA) <=210 MS AVERAGE V-CYCLE BETWEEN (B)(6) 2010 07:24:51 AND (B)(6) 2010 07:19:32. TWO - VF=190 MS ON (B)(6) 2010 12:21:50 AND ON (B)(6) 2010 08:28:09. VENTRICULAR SHORT INTERVAL COUNT V-SIC=97.4 COUNTS AVG/DAY, IN 2.10 DAYS, BETWEEN (B)(6) 2010 06:02:31 AND (B)(6) 2010 08:33:40.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD PRODUCED NOISE AND OVERSENSING. AFTER FURTHER INVESTIGATION, IT WAS DETERMINED THE LEAD HAD A FRACTURE. THE LEAD IS NO LONGER IN SERVICE AND WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD PRODUCED NOISE AND OVERSENSING. AFTER FURTHER INVESTIGATION, IT WAS DETERMINED THE LEAD HAD A FRACTURE. THE LEAD IS NO LONGER IN SERVICE AND WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT QUATTRO SECURE | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6947 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Hospitalization| O| R | (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4439 COMPETITOR IMPLANTABLE PACING LEAD| 4439 COMPETITOR IMPLANTABLE PACING LEAD| (B)(4) IMPLANTABLE PACEMAKER/CARDIO/DEFIB |