FDA Adverse Event Malfunction Summary report: N

STERLING SL BALLOON DILATATION CATHETER

MDR report key: 1990041 · Received February 15, 2011

Report

Report Number
2134265-2011-00279
Event Type
Malfunction
Date Received
February 15, 2011
Date of Event
January 18, 2011
Report Date
January 21, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LIT
PMA / PMN Number
K093720
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: (B)(6). DEVICE EVALUATED BY MANUFACTURER: MAGNIFIED INSPECTION REVEALED A HOLE IN THE SHAFT 69 CM FROM THE STRAIN RELIEF AND KINKS IN THE BALLOON AND INNER SHAFT LOCATED 8 AND 9.5 CM FROM THE TIP. THERE WAS NO EVIDENCE OF ANY MATERIAL OR MANUFACTURING DEFICIENCIES THAT COULD HAVE CONTRIBUTED TO THE DAMAGE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY TREATMENT PROCEDURE, A BALLOON RUPTURE OCCURRED. THE 100% CHRONIC TOTAL OCCLUSION WAS LOCATED IN A MODERATELY TORTUOUS AND SEVERELY CALCIFIED RIGHT POSTERIOR TIBIAL ARTERY. THE LESION WAS PRE-DILATED WITH A 1.5 X 20MM STERLING ES OVER-THE-WIRE BALLOON CATHETER. POST DILATATION WAS PERFORMED WITH THE 2.0MM X 120 X 150 STERLING SL MONORAIL BALLOON CATHETER. DURING THE FIRST INFLATION, THE BALLOON RUPTURED AT 3ATMS. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STERLING SL BALLOON DILATATION CATHETER CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT BOSTON SCIENTIFIC - MAPLE GROVE H74939147201210 13743925

Patients

Seq Age Sex Outcome Treatment
1 5FR 55CM MEDIKIT INTRODUCER SHEATH| 1.5 X 20MM STERLING ES| ASTATO XS GUIDE WIRE| CRUISE GUIDE WIRE