FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE¿

MDR report key: 19898034 · Received August 2, 2024

Report

Report Number
2024168-2024-09168
Event Type
Injury
Date Received
August 2, 2024
Date of Event
July 1, 2024
Report Date
October 8, 2024
Manufacturer
ABBOTT VASCULAR INC.
Product Code
MGB
UDI-DI
08717648113154
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

VISUAL INSPECTIONS WERE PERFORMED ON THE RETURNED DEVICE. THE REPORTED NEEDLE TO CUFF MISS WAS OBSERVED AS A LINK DETACHMENT. PRODUCTION RECORD AND CORRECTIVE AND PREVENTIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. THE REPORTED DIFFICULTIES AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR LINK PULLED UNTIL IT BREAKS DUE TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE REPORTED INFORMATION AND RESULTS OF THE COMPLAINT INVESTIGATION THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. D4 CORRECTION: LOT # UPDATED FROM 3091141 TO 3110741. H6 CORRECTION: REPORTED DEVICE CODE 2983 REMOVED.

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT.

Description of Event or Problem · 0

SUBSEQUENT TO THE INITIALLY FILED REPORT, THE FOLLOWING INFORMATION WAS RECEIVED: IT WAS DETERMINED THAT THE REPORTED MALFUNCTION WAS REFERRING TO A CUFF MISS [SUTURE RETRIEVAL ISSUE] AND NOT A PLUNGER JAM. NO ADDITIONAL INFORMATION WAS PROVIDED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF AN UNSPECIFIED ARTERY USING THE PRE-CLOSE TECHNIQUE VIA PRIOR TO A TRANSCATHETER AORTIC VALVE IMPLANTATION (TAVI) INTERVENTIONAL PROCEDURE. REPORTEDLY, THE SUTURE MECHANISM WHEN ACTIVATED DID NOT PLACE THE STITCH IN THE ARTERY WALL DURING STEP 2 [PLUNGER JAM]. THE SUTURE OF A NEW DEVICE WAS SUCCESSFULLY PRE-PLACED. THE TAVI PROCEDURE WAS COMPLETED. HEMOSTASIS WAS ACHIEVED WITH THE PRE-PLACED SUTURE. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1037575 PERCLOSE PROGLIDE¿ DEVICE, HEMOSTASIS, VASCULAR MGB ABBOTT VASCULAR INC. 12673-05 3110741 08717648113154

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention