TRUE METRIX
Report
- Report Number
- 1000113657-2024-00319
- Event Type
- Malfunction
- Date Received
- August 2, 2024
- Date of Event
- July 8, 2024
- Report Date
- August 28, 2024
- Manufacturer
- TRIVIDIA HEALTH INC
- Product Code
- NBW
- PMA / PMN Number
- K140100
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
SECTIONS WITH ADDITIONAL INFORMATION AS OF 28-AUG-2024: METER WAS RETURNED FOR EVALUATION TEST STRIPS WERE NOT RETURNED FOR EVALUATION. REPORTED DEFECT NOT REPRODUCED ON RETURNED METER. PRODUCT EVALUATION HAS BEEN COMPLETED. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-003: INCONSISTENT TEST METHOD.
INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER AND TEST STRIPS WERE NOT RETURNED FOR EVALUATION. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIP LOT TESTED WITHIN SPECIFICATIONS. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-003: INCONSISTENT TEST METHOD. NOTE: MANUFACTURER MADE SEVERAL ATTEMPTS TO CONTACT CUSTOMER TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - UNABLE TO ESTABLISH CONTACT WITH CUSTOMER.
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS AND ERROR MESSAGE (E-5). THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 140, 150, 166, 153 AND 258 MG/DL. THE CUSTOMER¿S EXPECTED BLOOD GLUCOSE TEST RESULT RANGE IS 124-127 MG/DL AM FASTING. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 10/17/2025 AND OPEN VIAL DATE IS 06/2024. THE METER MEMORY WAS NOT REVIEWED FOR PREVIOUS TEST RESULT HISTORY (CUSTOMER PROVIDED RESULTS FROM HER NOTES): TUESDAY ON (B)(6) 2024 ,140MG/DL, AM FASTING . WEDNESDAY ON (B)(6) 2024 ,150MG/DL, AM FASTING. THURSDAY ON (B)(6) 2024 ,166MG/DL, AM FASTING . FRIDAY ON (B)(6) 2024, 153MG/DL, AM FASTING . TODAY ON (B)(6) 2024, 258MG/DL ,AM FASTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 885522 | TRUE METRIX | SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER | NBW | TRIVIDIA HEALTH INC | STRIP, TRUE METRIX NFRS24/CS 50CT MG/DL | ZC5625S |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |