FDA Adverse Event Malfunction Summary report: N

NATURA

MDR report key: 19891481 · Received August 2, 2024

Report

Report Number
9618003-2024-02706
Event Type
Malfunction
Date Received
August 2, 2024
Date of Event
July 10, 2024
Report Date
July 10, 2024
Manufacturer
CONVATEC DOMINICAN REPUBLIC INC
Product Code
EXE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

DEVICE 9 OF 10. E1: COMPLAINANT CITY: (B)(6). COMPLAINANT STATE/PROVINCE: (B)(6). COMPLAINANT PHONE: (B)(6). BASED ON THE AVAILABLE INFORMATION, THIS EVENT IS DEEMED TO BE A REPORTABLE MALFUNCTION. PHOTOGRAPH, VIDEO AND/OR PHYSICAL SAMPLE EVALUATION: THERE ARE PHOTOGRAPHS ASSOCIATED WITH THIS CASE AND IN THESE, THE REPORTED DEFECT CAN BE SEEN. NO UNUSED RETURN SAMPLE WAS EXPECTED. BATCH RECORD REVISION RESULTS: LOT 2H02765 WAS MANUFACTURED 23/SEP/2022, IN AUTOMATION TOOLING SYSTEMS (ATS) #2 LINE, WITH A TOTAL OF (B)(4). THE COMPLAINTS INVESTIGATOR PERFORMED A BATCH RECORD REVIEW ON 19/JUL/2024, TO VERIFY IF ALL THE APPLICABLE PROCEDURES WERE FOLLOWED AND NO ISSUES WERE FOUND; ALL THE COMPONENTS FOR ASSEMBLY WERE CORRECT PER BILL OF MATERIALS (BOM) AND ALL THE TOOLING INFORMATION DOCUMENTED WAS ALSO CORRECT, SYSTEM APPLICATION PRODUCT (SAP) MATERIAL IDENTIFICATION (ID) 1222274 AND MANUFACTURING ORDER (B)(4). REVIEW OF THE BATCH RECORD SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE MANUFACTURING PROCESS AND FINAL PRODUCT RELEASE HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DISCREPANCY RELATED TO THIS ISSUE WERE FOUND WITHIN THE DOCUMENTATION. HISTORICAL COMPLAINTS REVIEW: ON 19/JUL/2024, COMPLAINTS INVESTIGATOR RAN A QUERY IN DATABASE IN ORDER TO VERIFY THE COMPLAINTS REPORTED FOR THE 2H02765 LOT FOR THE MALFUNCTION ¿SKIN BARRIER STARTER HOLE IS DEFECTIVE (E.G MISALIGNMENT OR OFF CENTER), LEAKAGE MAY OCCUR (MOLDABLE ONLY)¿ DEFECT AND NO ADDITIONAL COMPLAINTS WERE IDENTIFIED. HISTORICAL NONCONFORMANCE REVIEW: ON 19/JUL/2024, COMPLAINTS INVESTIGATOR RAN A QUERY IN DATABASE LOOKING FOR ANY IN PROCESS NONCONFORMANCE / CORRECTIVE ACTION / PREVENTIVE ACTIONS (CAPA)(S) ASSOCIATED TO THE MALFUNCTION ¿SKIN BARRIER STARTER HOLE IS DEFECTIVE (E.G MISALIGNMENT OR OFF CENTER), LEAKAGE MAY OCCUR (MOLDABLE ONLY)¿ DEFECT FOR THE LOT NUMBER 2H02765 AND AS RESULT, NO NONCONFORMANCE / CAPA (S) FOR THIS MALFUNCTION WERE GENERATED DURING THE MANUFACTURING PROCESS OF THE REFERENCED LOT. CURRENT QUALITY CONTROLS: BASED ON THE PROCESS INSTRUCTION (PI), THE FOLLOWING TESTS ARE PERFORMED IN THE MANUFACTURING LINES, IN ORDER TO IDENTIFY THIS FAILURE MODE IN OUR MANUFACTURING PROCESS: ADHESIVE DISC ALIGNMENT TEST: FREQUENCY: CONTINUOUS VISUAL INSPECTION. SAMPLE QUANTITY: CONTINUOUS VISUAL INSPECTION. ACCEPTANCE CRITERIA: ACCEPT = 0 / REJECT = 1. DEFECT RATE ANALYSIS: THERE HAVE ONLY BEEN 10 DEFECTIVE PARTS CONFIRMED TO DATE FROM A LOT SIZE 37800 PRODUCTS. THIS REPRESENTS A DEFECT RATE OF ONLY 0.026%, WHICH IS WELL WITHIN AN APPROPRIATE ACCEPTABLE QUALITY LEVEL (AQL) FOR THIS DEFECT WHICH SHOULD 0.25% BASED ON OUR STANDARD OPERATING PROCEDURE (SOP). IN ADDITION, ALL OF THE IN-PROCESS TESTING ON THIS LOT DID NOT FIND A SINGLE DEFECTIVE UNIT, WHICH CONFIRMS THAT THE LOT IS UNLIKELY TO BREACH AN AQL OF 0.25. THIS ISSUE CERTAINLY APPEARS TO BE AN ISOLATED INCIDENT, BUT MORE IMPORTANTLY TO DATE, IT IS WELL WITHIN OUR ACCEPTED AQL LEVEL FOR THIS TYPE OF FAILURE MODE OR DEFECT. CONCLUSIONS: AS PHOTOGRAPHS WERE AVAILABLE FOR THIS COMPLAINT ISSUE, THEY WERE EVALUATED AND AS A RESULT, THE REPORTED MALFUNCTION FOR THE OBSERVED PRODUCT WAS CONFIRMED, HOWEVER, AFTER PERFORMING THE DEFECT RATE ANALYSIS FOR THIS ISSUE, IT WAS OBSERVED THAT THE QUANTITY OF REPORTEDLY AFFECTED PRODUCTS IS WITHIN THE APPROPRIATE AQL FOR THIS DEFECT. A BATCH RECORD REVIEW WAS COMPLETED AND SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE MANUFACTURING PROCESS AND FINAL PRODUCT RELEASE HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DISCREPANCY RELATED TO THIS ISSUE WERE FOUND WITHIN THE DOCUMENTATION. THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET PRODUCT MONITORING REVIEW PROCESS, STANDARD OPERATING PROCEDURE (SOP). TO DATE NO ADDITIONAL INFORMATION HAS BEEN RECEIVED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. FDA REGISTRATION NUMBER, REPORTING SITE: 1049092, MANUFACTURING SITE: 9618003.

Description of Event or Problem · 0

THE DISTRIBUTOR REPORTED THAT THE EDGE OF THE CENTRAL HOLE OF THE COMPANY'S KNOWN MOLDABLE WAFER WAS NOT REGULAR AND THE GLUE OVERFLOWED. THE PRODUCT WAS NOT USED BY THE END USER. PHOTOGRAPHS DEPICTING THE ISSUE WERE RECEIVED FROM THE COMPLAINANT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
983974 NATURA PROTECTOR, OSTOMY EXE CONVATEC DOMINICAN REPUBLIC INC 411803 2H02765

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown