FDA Adverse Event Injury Summary report: N

CONCERTO CRT-D DR

MDR report key: 1989032 · Received February 14, 2011

Report

Report Number
6000144-2011-00513
Event Type
Injury
Date Received
February 14, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S031
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE DEVICE IS IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. EVALUATION SUMMARY: (B)(4): WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. TIME OF ERI (ELECTIVE REPLACEMENT INDICATOR) IN SAVE TO DISK ON (B)(6) 2010 02:15:04, DEVICE RRT<=2.62 V. PATIENT ALERT FOR LOW BATTERY VOLTAGE ON (B)(6) 2010 02:15:04. WEEKLY BATTERY VOLTAGE TREND DATA SHOWS MIN BAT=2.64 TO 2.62 VOLTS BETWEEN (B)(6) 2010 AND (B)(6) 2010.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. TIME OF ERI (ELECTIVE REPLACEMENT INDICATOR) IN SAVE TO DISK ON (B)(6) 2010 02:15:04, DEVICE RRT<=2.62 V. PATIENT ALERT FOR LOW BATTERY VOLTAGE ON (B)(6) 2010 02:15:04. WEEKLY BATTERY VOLTAGE TREND DATA SHOWS MIN BAT=2.64 TO 2.62 VOLTS BETWEEN (B)(6) 2010. ANALYSIS OF THE DEVICE FOUND NORMAL BATTERY DEPLETION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD BATTERY VOLTAGE OF 2.9 V AND THEN ONE WEEK LATER WAS AT RRT (RECOMMENDED REPLACEMENT TIME). THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

ASKU

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONCERTO CRT-D DR IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. C154DWK ASKU

Patients

Seq Age Sex Outcome Treatment
1 52 YR Hospitalization| R (B)(4) TISSUE VALVE| 4194 IMPLANTABLE PACING LEAD| 4194 IMPLANTABLE PACING LEAD| (B)(4) TISSUE VALVE| 4592 IMPLANTABLE PACING LEAD| 4592 IMPLANTABLE PACING LEAD| 6947 IMPLANTABLE TACHY LEAD| 6947 IMPLANTABLE TACHY LEAD