BD NEEDLE PRECISIONGLIDE 16X1-1/2IN
Report
- Report Number
- 3003916417-2024-00185
- Event Type
- Malfunction
- Date Received
- August 1, 2024
- Date of Event
- July 17, 2024
- Report Date
- October 15, 2024
- Manufacturer
- BECTON DICKINSON IND. CIRURGICAS LTDA
- Product Code
- FMI
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- PHARMACIST
- Health Professional
- Yes
Narratives
B.3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.
SEVEN SAMPLES RECEIVED BY OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, DIRT ON THE SHIELD, DAMAGED SHIELD, EPOXY ON THE NEEDLE, AND LINT ON THE SHIELD ARE OBSERVED. EPOXY IS AN ADHESIVE USED TO JOIN THE CANNULA AND HUB. NO FOREIGN MATTER OBSERVED ON THE HUB, ADDITIONALLY, NO DEFECTS OR ISSUES OBSERVED ON THE PACKAGING OF THE PRODUCT. THERE ARE NO PUNCTURES OR HOLES. FUNCTIONAL TESTING WAS PERFORMED, NO CONTAMINATION WERE FOUND INSIDE THE PACKAGING. RETENTION SAMPLES FROM THE SAME LOT WERE EVALUATED, NO DEFECTS OR ISSUES OBSERVED. POSSIBLE ROOT CAUSE FOR DAMAGED SHIELD IS ASSOCIATED WITH PROTECTOR ENTANGLEMENT IN THE ASSEMBLY LINE BECOMING MISALIGNED CAUSING THE INCIDENT. LINT ON PROTECTOR IS ASSOCIATED WITH PLASTIC RESIDUE INSIDE THE CAVITY OF THE INJECTION POINT. NEW MOLDING REPLACEMENT WILL BE IMPLEMENTED. EPOXY AND DIRT ON THE SHIELD ARE ASSOCIATED WITH RESIN APPLICATION. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES.
IT WAS REPORTED THAT THE BD NEEDLE PRECISIONGLIDE 16X1-1/2IN HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER TRANSLATED FROM PORTUGUESE TO ENGLISH: PROBE 1.60X40MM (16GX1 1/2''). QUANTITY: (B)(4). LOT: 2243422/ REASON: 1 UNIT WITH A GELATINOUS FOREIGN BODY ON THE INSIDE OF THE PROTECTIVE CAP, 1 CRACKED PROTECTIVE CAP, 1 EXTENSION OF THE PROTECTIVE CAP PIERCING THE PLASTIC PACKAGING, 1 FOREIGN BODY IN THE BARREL, 1 FOREIGN BODY IN THE PROBE BODY, 2 BARRELS LEAKING GLUE FROM THE WELDS. ADDITIONAL INFORMATION PROVIDED: DATE OF THE EVENT? SEGREGATED DAILY DURING THE MONTHS OF MAY TO JULY/2024. TO COLLECT SAMPLES: PLEASE INFORM: ORGANIZATION NAME: ESSENTIAL DRUGS AND NUTRITION. CNPJ NUMBER: 20.511.326/0001-90. ICMS TAXPAYER (YES/NO): YES. PRODUCT PURCHASE INVOICE NUMBER: NF: (B)(4) (PURCHASE VIA SURGICAL MAFRA) NF: (B)(4) (PURCHASE VIA SURGICAL SANTA CRUZ) NF: (B)(4) (PURCHASE VIA SURGICAL SANTA CRUZ) NF: (B)(4) (PURCHASE VIA SURGICAL SANTA CRUZ). COMPLETE ADDRESS (STREET, ZIP CODE, NEIGHBORHOOD, CITY AND STATE); (B)(6) - PR. 85.802-030. SECTOR/DEPARTMENT: STOCK. CONTACT (NAME AND TELEPHONE) (B)(6). HOW MANY UNITS ARE AVAILABLE FOR COLLECTION? ALL UNITS NOTIFIED ((B)(4) UNITS). IS THE SAMPLE CONTAMINATED? IF SO, PROVIDE THE SUBSTANCE. NO.
NO ADDITIONAL INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 885979 | BD NEEDLE PRECISIONGLIDE 16X1-1/2IN | NEEDLE | FMI | BECTON DICKINSON IND. CIRURGICAS LTDA | 2243422 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |