MINIMED QUICK-SET
Report
- Report Number
- 3003442380-2024-18510
- Event Type
- Malfunction
- Date Received
- August 1, 2024
- Date of Event
- June 19, 2024
- Report Date
- November 25, 2025
- Manufacturer
- UNOMEDICAL A/S
- Product Code
- FPA
- PMA / PMN Number
- K160648
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. PER REVISION 21 OF PROCEDURE (B)(4), A CHILD INVESTIGATION IS NOT REQUIRED TO DOCUMENT THE DHR REVIEW FOR A TYPE 2 REPORTABLE COMPLAINT. HOWEVER, THE CHILD WAS CREATED TO ALLOW THE PARENT RECORD TO ADVANCE TO REVIEW AND SUMMARY. SINCE IT WAS CREATED, THE DHR REVIEW WAS DOCUMENTED WITHIN THE CHILD. COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6002453, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6002453 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 77 AND PACKAGING IN THE MACHINE MULTIVAC 12 ON 29-JUL-2023, WITH A TOTAL OF (B)(4) UNITS. ASSEMBLY: THE SUB-ASSEMBLY, ASSEMBLY OF THE LOT 3G03307 WAS MANUFACTURED ACCORDING TO THE WI VERSION 26 AND MANUFACTURED IN THE LINE ASSEMBLY OF QUICK SET, ON 28-JUL-2023, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, ASSEMBLY OF THE LOT 3G03308 WAS MANUFACTURED ACCORDING TO THE WI VERSION 26 AND MANUFACTURED IN THE LINE ASSEMBLY OF QUICK SET, ON 29-JUL-2023, WITH A TOTAL OF (B)(4) UNITS. GLUING TUBE: THE SUB-ASSEMBLY, GLUING TUBE OF THE LOT 3G03009 WAS MANUFACTURED ACCORDING TO THE WI VERSION 38 AND MANUFACTURED IN THE MACHINE GLUING 05, ON 27-JUL-2023, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING TUBE OF THE LOT 3G02994 WAS MANUFACTURED ACCORDING TO THE WI VERSION 38 AND MANUFACTURED IN THE MACHINE GLUING 04, ON 27-JUL-2023, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING TUBE OF THE LOT 3G03007 WAS MANUFACTURED ACCORDING TO THE WI VERSION 38 AND MANUFACTURED IN THE MACHINE GLUING 08, ON 28-JUL-2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT AN EXTENDED INSPECTION WAS CREATED DUE TO LEAK FAILURE, EXTENDED INSPECTION WAS FOUND WITHIN SPECIFICATION. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE; NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. EXTENDED INSPECTION IS NOT RELATED WITH THE CLAIMED MALFUNCTION. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
E1: PATIENT CITY: (B)(6). PATIENT COUNTRY: UNITED ARAB EMIRATES.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED ARAB EMIRATES. ON 19-JUN-2024, IT WAS REPORTED THAT THE PATIENT FACED LEAKING AT INFUSION SET TUBING WHICH FURTHER CAUSE THAT TUBING CAME APART AT THIGH. THE INFUSION SET WAS IN USE FOR ONE DAY. THE PATIENT REPORTED THAT THERE WAS NOT STRESS OR PULL ON THE TUBING. NO FURTHER INFORMATION AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 433199 | MINIMED QUICK-SET | UNO QUICK-SET 60/6 SC1 MECA | FPA | UNOMEDICAL A/S | MMT-399A | 6002453 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |