FDA Adverse Event Malfunction Summary report: N

CD HORIZON® SPINAL SYSTEM

MDR report key: 19879958 · Received August 1, 2024

Report

Report Number
1030489-2024-00906
Event Type
Malfunction
Date Received
August 1, 2024
Date of Event
July 10, 2024
Report Date
September 23, 2024
Manufacturer
MEDTRONIC SOFAMOR DANEK USA, INC
Product Code
HXX
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS OF PART# 5484113, LOT# K22L1403 - VISUAL AND OPTICAL EXAMINATION IDENTIFIED THAT THE TIP OF THE DRIVER IS STARTING TO STRIP. THE CRISP EDGES NEAR THE TIP HAVE BEEN ROLLED OVER. THIS APPEARS TO BE THE RESULT OF WEAR AND NOT A FAILURE AFTER A SINGULAR USE. OBSERVATIONS ARE CONSISTENT WITH WEAR AFTER REPEATED USE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL VIA A MANUFACTURER REPRESENTATIVE REGARDING THE INSTRUMENTS USED FOR POSTERIOR SPINAL FUSION THERAPY. IT WAS REPORTED THAT BOTH OF THE TWISTED DRIVER TIPS ARE EITHER WORN OR BROKEN OFF. THERE WAS NO PATIENT INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1188712 CD HORIZON® SPINAL SYSTEM SCREWDRIVER HXX MEDTRONIC SOFAMOR DANEK USA, INC 5484113 K22L1403

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown