FDA Adverse Event Malfunction Summary report: N

BD Q-SYTE EXT SET

MDR report key: 19877821 · Received July 31, 2024

Report

Report Number
1710034-2024-00805
Event Type
Malfunction
Date Received
July 31, 2024
Date of Event
June 30, 2024
Report Date
August 9, 2024
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FPA
PMA / PMN Number
K013621
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 385102 AND LOT NUMBER 1210970. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED.

Additional Manufacturer Narrative · 0

H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD Q-SYTE EXT SET LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: ON (B)(6) 2024, THE CHILD NEEDED REHYDRATION TREATMENT DUE TO HIS CONDITION. AT 13:00, INTRAVENOUS INFUSION WAS GIVEN AND A NEEDLELESS SEALED INFUSION CONNECTOR WAS CONNECTED. WHEN INSPECTING THE PATIENT AT 13:30, THE BED UNIT WAS FOUND TO BE DAMP. AFTER INVESTIGATION, IT WAS FOUND THAT THE NEEDLELESS SEALED INFUSION CONNECTOR WAS IF THERE IS WATER LEAKAGE, REPLACE THE CONNECTOR WITH A NEW ONE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
739890 BD Q-SYTE EXT SET SET, ADMINISTRATION, INTRAVASCULAR FPA BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 1210970

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown