FDA Adverse Event Malfunction Summary report: N

BEACH CHAIR POSITIONER

MDR report key: 19867989 · Received July 30, 2024

Report

Report Number
3010216206-2024-00009
Event Type
Malfunction
Date Received
July 30, 2024
Date of Event
June 18, 2024
Report Date
October 2, 2024
Manufacturer
ALLEN MEDICAL SYSTEMS
Product Code
CCX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE BEACH CHAIR SHOULDER POSITIONER IS DESIGNED TO POSITION AND SUPPORT THE PATIENT¿S HEAD AND TORSO IN A VARIETY OF SURGICAL PROCEDURES INCLUDING, BUT NOT LIMITED TO ORTHOPEDIC SURGERIES LIKE, ROTATOR CUFF, TOTAL SHOULDER, REVERSE TOTAL SHOULDER, SLAP REPAIR AND OTHER SHOULDER SURGERY. THESE DEVICES ARE INTENDED TO BE USED BY HEALTHCARE PROFESSIONALS WITHIN THE OPERATING ROOM SETTING. BAXTER IS IN THE PROCESS OF INSPECTING THE AR-1627 BEACH CHAIR POSITIONER. THERE WAS NO ALLEGATION OF PATIENT OR CAREGIVER INJURY OR DEATH REPORTED FROM THIS ALLEGED INCIDENT. INVESTIGATION IS ONGOING. ALL ADDITIONAL AND RELEVANT INFORMATION THAT IS IDENTIFIED FOLLOWING COMPLETION OF THE INVESTIGATION WILL BE SUBMITTED IN A FINAL REPORT.

Additional Manufacturer Narrative · 0

THE BEACH CHAIR SHOULDER POSITIONER IS DESIGNED TO POSITION AND SUPPORT THE PATIENT¿S HEAD AND TORSO IN A VARIETY OF SURGICAL PROCEDURES INCLUDING, BUT NOT LIMITED TO ORTHOPEDIC SURGERIES LIKE, ROTATOR CUFF, TOTAL SHOULDER, REVERSE TOTAL SHOULDER, SLAP REPAIR AND OTHER SHOULDER SURGERY. THESE DEVICES ARE INTENDED TO BE USED BY HEALTHCARE PROFESSIONALS WITHIN THE OPERATING ROOM SETTING. THE BEACH CHAIR SHOULDER POSITIONER WAS RECEIVED BY BAXTER. UPON INSPECTION, IT WAS DETERMINED THAT THE KNOBS TO LOCK TO THE FIXTURE WERE DAMAGED CAUSING THE RUPTURE OF THE BEACH CHAIR BACKBOARD . THE 4 LOBE BLACK KNOBS TO LOCK TO THE FIXTURE, BEACH CHAIR BACKBOARD, VELCRO AND SCREW COVERS WERE REPLACED. BASED ON THIS INFORMATION, NO FURTHER ACTIONS ARE REQUIRED AT THIS TIME. ALTHOUGH THERE WAS NO REPORTED INJURY WITH THIS EVENT, IF THE REPORT OF A BROKEN BEACH CHAIR POSITIONER BACK PANEL WERE TO RECUR, IT COULD POTENTIALLY CAUSE SERIOUS INJURY OR DEATH. THEREFORE BAXTER IS REPORTING THIS EVENT.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THAT DURING A TOTAL SHOULDER ARTHROPLASTY PROCEDURE, AN AR-1627 BEACH CHAIR POSITIONER WAS BROKEN ON THE PANEL ON THE BACK, LEADING THE HYDRAULIC NOT TO WORK AND COMPLETELY COLLAPSE. NO HARM OR SIGNIFICANT IMPACT TO THE SURGICAL PROCEDURE WAS REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

Description of Event or Problem · 0

THE CUSTOMER REPORTED THAT DURING A TOTAL SHOULDER ARTHROPLASTY PROCEDURE, AN AR-1627 BEACH CHAIR POSITIONER WAS BROKEN ON THE PANEL ON THE BACK, LEADING THE HYDRAULIC NOT TO WORK AND COMPLETELY COLLAPSE. NO HARM OR SIGNIFICANT IMPACT TO THE SURGICAL PROCEDURE WAS REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
215579 BEACH CHAIR POSITIONER SUPPORT, PATIENT POSITION CCX ALLEN MEDICAL SYSTEMS AR-1627

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown