FDA Adverse Event
Malfunction
Summary report: N
PHOTOTHERAPY SYSTEM 7850
MDR report key: 198624
·
Received November 18, 1998
Report
- Report Number
- 2510954-1998-00007
- Event Type
- Malfunction
- Date Received
- November 18, 1998
- Date of Event
- October 16, 1998
- Report Date
- October 16, 1998
- Manufacturer
- HILL-ROM AIR-SHIELDS
- Product Code
- LBI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
THE CUSTOMER REPLACED THE PHOTO THERAPY LAMPS WITH INCORRECT LAMPS, NOT PROVIED BY HILL-ROM AIR-SHIELDS. THE RESULT WAS SHATTERING OF THE INFRARED (IR) FILTER, WHICH FELL ON TO THE ULTRAVIOLET (UV) FILTER CAUSING IT TO PARTIALLY MELT. NO INJURY OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PHOTOTHERAPY SYSTEM 7850 | NEONATAL PHOTOTHERAPY | LBI | HILL-ROM AIR-SHIELDS | PTM78 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN |